Section 11: Utilities
164
• From the Main Menu press Battery Management.
• Press
Charger Control.
• Press
to begin charging. A flashing 'CHARGING'
message will be displayed.
If there is no power coming from the power transformer a flashing
'NO CHARGER' message will be displayed. If this happens, check
that the mains power is turned on and the battery adapter is
connected properly.
Charging stops automatically when complete or you can press
in the Charger Control Menu to stop the charging. A full charging
cycle takes approximately 2 1/2 hours.
Reconditioning the Battery Pack
Warning: Read the warnings in Battery Management (page 161)
before performing this operation.
Battery reconditioning is the process of first discharging and then
recharging batteries. It reduces the undesirable memory effect in the
Ni-Cad battery. The battery pack should be reconditioned the first
time it is used and at least once a month thereafter to prevent battery
degradation. The recondition process will take up to two hours,
depending on initial battery charge.
Caution: If the battery pack is not reconditioned the first time it is
used, the battery charge displayed may be incorrect.
• Plug the instrument power transformer to a standard power
outlet and its connector to the instrument charger socket.
• From the Main Menu press
Battery Management.
• Press
Charger Control.