Section 2: Basic Operation
22
magnetic accelerometer base or via a threaded stud on the
machine surface. The accelerometer should not move
independently of the machine part it is attached to.
Oriented correctly as vibration can differ greatly with respect
to direction.
Attached to exactly the same position as before if this is an
ongoing study of a particular measurement point (mark the
position if necessary).
Clear from other cables and not twisted, kinked or tangled.
Permanent Mounting
To mount using the stud method, prepare a mounting spot on the
machine following the specifications shown in the diagram.
Unscrew the magnetic base from the accelerometer and
screw onto the threaded stud. Use of thread locking
compound is suggested.
Stud Mounting Spot
Caution: Do not use a wrench to tighten the accelerometer as this
will damage the sensor.