EasyManua.ls Logo

Conair Thermolator TW-1 - System Testing Procedures

Conair Thermolator TW-1
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Conair Thermolators with TW-1 or TW-2 controls provide a
Test Mode that tests displays and keys on the control panel, as
well as inputs and outputs. The Test Mode also allows calibra-
tion of the supply and return line thermocouples.
System tests and calibration should be performed annually.
To enable Test Mode:
Press to shut down the Thermolator.
Disconnect and lock out main power to the unit,
then open the electrical enclosure.
Set dip switches 3 and
5 to OFF.
Set dip switch 6 to ON.
Close the electrical enclosure and restore
power to the unit.
u The control displays “tSt.”
u The Test Mode LED lights.
Press any button to display the first test menu.
Test mode provides the following menus:
l Key/Display Test
l Input Test
l Output Test
l Calibration
l Total Operating Hours
l Output Monitor Enable/Disable
After performing each test, hold the Setpoint
for 3 seconds to index to the next test menu. The proce-
dure for each test is described on the following pages.
TW-1, TW-2
C
ONTROL
PERFORMING
SYSTEM TESTS
UGH025-0504 Thermolator TW Series
MAINTENANCE 5-5
Stop
876
54
3
21
1
2
3
4
5
6
7
IMPORTANT: All normal
operating functions are
disabled while Test Mode is
enabled. To return to normal
operation, you must disable
Test Mode.
To disable Test Mode,
repeat steps 2 through 5,
setting dip switch 6 to OFF
instead of ON.

Table of Contents

Related product manuals