INTENDED USE
The ConMed Linvatec Endoscopes are indicated for use when access,
illumination, and/or observation of body cavities, hollow organs, and/or
canals is necessary during minimally invasive surgical procedures.
CONTRAINDICATIONS
The ConMed Linvatec Endoscopes are contraindicated in situations where
endoscopic techniques and minimally invasive surgery are contraindicated.
WARNINGS
1. It is important to read all instructions carefully. Failure to follow
warnings and/or cautions may lead to improper use of the
endoscope that can lead to severe patient injury and/or damage to the
endoscope.
2. It is recommended that only individuals properly trained in surgical
endoscopy use the endoscope and only in approved endoscopic
surgical procedures.
3. To prevent possible burns to patient tissue, do not leave the
endoscope distal tip in direct contact with patient tissue.
4. The fiber optic light post may become hot to touch during use. Avoid
hand or tissue contact with the light post and/or reduce the light source
brightness setting to avoid high temperatures.
PRECAUTIONS
1. ConMed Linvatec endoscopes are not to be bent or flexed. This
may cause damage or vastly reduce the endoscope’s life.
Please follow the recommended steps outlined below.
2. Use extreme care when handling the endoscope.
3. Carefully clean the exposed lenses and fiber optic post with a cotton
swab and isopropyl alcohol after each use.
4. Do not allow hard objects to come into contact with the objective or
ocular lenses.
5. Do not use the endoscope for leverage during surgery.
INSTRUCTIONS FOR USE
1. Sterilize endoscopes prior to use.
2. Always handle the endoscope by grasping the eyepiece. Handling by
the shaft may damage the endoscope.
3. Prior to and after use, visually inspect the ocular lens (proximal end),
objective lens (distal end) and the fiber optic post for damage or buildup
of deposits (if deposits exist, please see “Removing Deposits” for
cleaning instructions).
4. Avoid contact of the objective lens (distal end) with any
instrumentation, especially resection systems.
5. Avoid touching the lenses. Debris and fingerprints obscure the view.
Components
Your ConMed Linvatec endoscope comes
complete with the following endoscope
adapters: ACMI, WOLF DYONICS and
LINVATEC/STORZ.
Endoscopic Assembly:
1. Objective lens (distal lens)
2. Rod lens tubular shaft
3. Ocular lens (proximal lens) with
eyepiece
4. Fiber optic post (ACMI style)
5. Fiber optic post adapter (WOLF, DYONICS style)
6. Fiber optic post adapter (LINVATEC/STORZ style)
1
2
6
5
4
3