What does '[al' on ConMed SYSTEM 5000 Medical Equipment display mean?
- JJennifer BlakeSep 13, 2025
The '[al' message on your ConMed Medical Equipment display indicates that the system is in Calibration Mode.
What does '[al' on ConMed SYSTEM 5000 Medical Equipment display mean?
The '[al' message on your ConMed Medical Equipment display indicates that the system is in Calibration Mode.
Details the two-year warranty, contact numbers for service, and legal information.
Explains key functional parameters for each operational mode, including Cut, Coag, and Bipolar modes.
Describes how mains power is converted to electrosurgical output through various system components.
Details configurable settings via dipswitches on the controller board, affecting system behavior.
Recommends periodic inspection and performance testing to ensure safe and effective operation.
Specifies that only qualified biomedical engineers should perform service on the unit.
Provides instructions for unit disassembly and reassembly, including cautions about static electricity.
Details how to clean the interior and exterior of the unit using appropriate methods and materials.
Outlines checks for power cord, connectors, external damage, and debris at least every six months.
Instructions to measure resistance between power plug ground and equipotential ground connection.
Procedure to verify proper operation of displays, alarms, and commands as per Section 2.3.1.
Details equipment and procedures for testing monopolar and bipolar output power accuracy across various loads.
Explains RF leakage hazards and presents allowable leakage currents to ground for various configurations.
Details procedures and limits for measuring line frequency leakage currents to neutral and ground.
Verifies correct operation of the A.R.M. circuit by testing its response to known resistances.
Details how to check output coupling capacitor values for Monopolar and Bipolar modes.
Explains how to enter calibration mode and the initial display indications for calibration status.
Details how to select modes and perform calibration for Monopolar, Bipolar, and A.R.M.™ systems.
Provides procedures for accessing stored error codes and clearing the fault memory.
Explains how to view system configuration DIP switch settings without removing the top cover.
Details how to use DACview in Test Mode to output internal system readings to a DVM or oscilloscope.
Lists potential errors, their causes, and recommended troubleshooting steps or corrections.
Addresses issues like no output power, low voltage supply problems, and MOSFET failure.
Emphasizes safety when working with high voltage, including using isolation transformers.
Provides guidance on checking LEDs, fuses, gate drive signals, and feedback resistors for HVPS issues.
Lists required information for ordering parts and obtaining return authorization from CONMED.
Lists numeric fault codes, their descriptions, possible causes, checks, and corrections.
Continues the list of fault codes, providing further diagnostic information and corrective actions.
| Type | Electrosurgical Generator |
|---|---|
| Display | Digital |
| Input Voltage | 100-240 VAC |
| Line Frequency | 50/60 Hz |
| Modes | Cut, Coag |
| Safety Features | Return Electrode Monitoring (REM) |
| Footswitch | Optional |