Service Guide SD One+ 71
Error Code:
100-50231
Keypad Error Code:
BOOT (1)
Level:
RD Correctable
Description:
Firmware download is in progress
Potential Failed Component:
Firmware
Corrective Action:
Firmware.
Error Code:
100-50232
Keypad Error Code:
ERR_GENERAL (3)
Level:
User Correctable
Description:
Firmware is incomplete. Please download new firmware.
Potential Failed Component:
User.
Corrective Action:
User interaction.
Error Code:
100-50234
Keypad Error Code:
None.
Level:
User Correctable
Description:
Scanner is in safemode. Please reboot scanner or download new firmware.
Potential Failed Component:
User, SUx
Corrective Action:
User interaction.
Main controller board (SUx).
Error Code:
100-50237
Keypad Error Code:
ERR_GENERAL (3)
Level:
User Correctable
Description:
Host PC does not support SSE2 calculations.
Potential Failed Component:
User.
Corrective Action:
User interaction.
Error Code:
100-20088
Keypad Error Code:
ERR_GENERAL (3)
Level:
Support Correctable
Description:
Motor system error, motor movement not detected.
Potential Failed Component:
Mech, SUx, Cable, Motor
Corrective Action: