CONTOUR NEXT GEN - TECHNICAL INFORMATION: SYSTEM ACCURACY; System Accuracy Results Comparison; User Accuracy Study Findings

CONTOUR NEXT GEN
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
64 65
Contour
®
next GEN blood glucose monitoring system
TECHNICAL INFORMATION
Table 2: System accuracy results for glucose concentration
≥ 5.55 mmol/L
Dierence range in values
between YSI laboratory
reference method and
Contour next GEN meter
Within
± 5%
Within
± 10%
Within
± 15%
Number (and percent) of samples
within specied range
262 of
408
(64.2%)
397 of
408
(97.3%)
408 of
408
(100%)
Table 3: System accuracy results for glucose concentrations from
1.9 mmol/L to 29.1 mmol/L
Within ± 0.83 mmol/L or ± 15%
600 of 600 (100%)
Acceptance criterion in ISO 15197:2013 is that 95% of the measured
glucose values shall fall within either ± 0.83 mmol/L of the average
measured values of the reference measurement procedure at
glucose concentrations < 5.55 mmol/L or within ± 15% at glucose
concentrations ≥ 5.55 mmol/L.
User Accuracy
A study evaluating glucose values from ngertip capillary blood
samples obtained by 324 lay persons showed the following results:
100% within ± 0.83 mmol/L of the medical laboratory values at glucose
concentrations < 5.55 mmol/L and 98.60% within ± 15% of the medical
laboratory glucose concentrations ≥ 5.55 mmol/L.
User Accuracy Radar Plot
1.7
mmol/L
>1.9
mmol/L
1.1
mmol/L
1.4
mmol/L
0.8
mmol/L
0.6
mmol/L
0.3
mmol/L
2
0
.
0
3
3
.
3
0
.
8
1
6
.
1
2
0
.
0
1
6
.
1
1
1
.
1
2
3
.
7
2
7
.
8
Y
S
I
®
(
m
m
o
l
/
L
)
5
.
6
2
3
.
7
1
1
.
1
5
.
6
2
7
.
8
10% 20% 30% >35%5% 15% 25%
+
All samples (n = 324) 0.8 mmol/L or 15%0.6 mmol/L or 10%
The 324 data points represent the blood glucose monitoring system
(BGMS) results. The distance from the center represents the
magnitude of the error compared with the laboratory reference result.