EasyManua.ls Logo

CONTROL SOLUTIONS LogTag VFC400-USB - Interpreting Todays Data

CONTROL SOLUTIONS LogTag VFC400-USB
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reviewing Day Statistics on the Display VFC400-USB User Guide 44
Today's data
Pressing the button displays the current day’s
maximum statistic, in this case, the review starts
with channel 1:
l
The Today marker flashes to indicate that
today's data is being displayed.
l
Today, no temperature values were recorded
above the upper alarm threshold, and no
alarm event was generated.
l
The maximum temperature recorded today was
4.8°C.
l
The paused function is enabled.
Pressing the button again displays the current
day’s minimum statistic:
l
still flashes, as the same day's data (today)
is being displayed.
l
Today, no temperature values were recorded
below the lower alarm threshold, and no alarm
event was generated.
l
The minimum temperature recorded today was
1.4°C.
Yesterday's data
Pressing the button now displays yesterday’s
maximum statistic:
l
The -1 marker flashes to indicate that
yesterday's data is being displayed.
l
Yesterday, recordings above the upper
threshold were recorded, indicated by the
upper alarm threshold marker .
l
The duration value shows the amount of time
recorded above the limit yesterday, which was
12 hours 30 minutes.
l
An alarm was triggered, indicated by the , as
this duration was longer than the allowed time
above the threshold.
l
The maximum temperature recorded yesterday
was 12.9°C.