Service manual OEB/OES/OGB/OGS
KD / 31.03.2009 5_00d_Fehlersuchliste_h_En.doc
Page 16 / 21
OEB
OES
OGB
OGS
Defect
Possible cause or remedy
5.4 Condenser
5.4.1 Constant water supply in
the condenser
Solenoid valve Y1 is not closing
a) Solenoid valve contaminated
b) Solenoid valve defective
c) Water pressure too low
Short circuit of the B3 sensor wiring
Appliance connected to hot water Cold
water
Solenoid valve Y1 reversed with Y2 or Y3
5.4.2 No supply of cooling water
in the condenser or supply too
low
Check polarity of the B3 sensor connection
(green/white)
Water supply pipe closed
Condenser quenching nozzle contaminated
Dirt filter in hose fitting contaminated
Inlet filter in the solenoid valve
contaminated Clean
Solenoid valve Y1 is not opening
a) Solenoid coil defective
b) Solenoid valve defective
Short circuit of the B3 - sensor wiring or
cooling sensor B3 defective
5.5 Motor
5.5.1 Motor does not run after
start of programme
Appliance is set to dummy version
Door contact switch defective
Twist lever catch (catch screw) too far from
the Reed contact
Test door contact switch in service
programme r23; if necessary, make
appliance level (above all with floor models)
Motor contactor K5/K50/K55/K56/K57 does
not pull in
a) Contact problem on the terminals
b) Motor contactor defective
Fan malfunction (excessive temperature) E
03
Test motor contactors (K5, K50, K55, K56,
K57) for sticking contacts