The Freedom
TM
CPAP Battery Kit comes with the items listed below.
The following items are required to connect certain PAP devices to the
Freedom™ battery. These items are typically packaged separately and are
not included in the Freedom
TM
CPAP Battery Kit box.
FREEDOM
TM
CPAP BATTERY AC POWER SUPPLY DC POWER PIGTAIL CORD
CONNECTOR TIPS (3) DC INPUT CORD
1
RESMED
24V POWER CONVERTER
{ITEM #: BPS C-24V-RES}
DREAMSTATION
12V POWER ADAPTER
{ITEM #: BPS C-12V-DS}
150W SINE WAVE
POWER INVERTER
{ITEM #: BPS INVERTER}
Required for ResMed
Air 10 and S9 devices.
Required for Respironics
DreamStation devices.
Required for most all
other PAP device models.
Package Contents
Connection Components Sold Separately
34
FAQ | Troubleshooting
Below are the most common questions/issues regarding use of the
Freedom
TM
CPAP Battery. If you have additional questions please contact
us at 877-445-5228 or email us at info@batterypowersolutions.net.
LOWER THAN EXPECTED RUN TIMES
Several factors can impact how long the battery will power a PAP device.
Below are the most common of these factors:
The use of heated humidification or heated tubing will reduce run times
by two thirds in most cases and in some cases could prevent the battery
from being able to power the PAP device. You should disconnect the
humidifier or completely turn it off and use non-heated tubing to obtain
optimal battery performance.
If humidification is required, try lowering the humidification setting as
many people can lower their heat setting and still achieve the desired
result. If lowering the setting does not produce enough of an increase
in run time you can use the battery in the dual battery configuration
to increase the run time.
Very high pressure settings (above 18 cm/H2O) can impact run times
especially when other factors are present. All average run time testing
is done on a pressure setting of 10 cm/H2O.
Leaking masks can impact run times as the PAP device will be working
harder to maintain the prescribed pressure setting and thus be drawing
more power from the battery.
BiLevel devices will have lower run times as the pressure setting is
not constant and changes throughout the night requiring more power
to operate the device.
Some older obsolete PAP device models and newer targeted ventilation
PAP devices require a much higher power draw than todays standard
CPAP devices and therefore can result in lower run times or could
prevent the battery from being able to power the device.
The device-specific connection components required for some PAP
device models do draw power from the battery but this has been
factored into the listed average run times.
NOTE: The included DC power pigtail cord and connector tips included in
the battery kit are compatible with Respironics System One, Respironics M
series and DeVilbiss IntelliPAP 1 series devices. All other PAP device makes
and models will require an additional connection component (see below).
{Letters: N, W, J}