Unpacking
Your T4 will have been inspected and quality checked before it left our manufacturing facility.
It will be congured as a standard unit with standard settings as shown in the table below and
any changes to suit your specic site requirements can be made utilising Portables Pro 2.0 PC
Application and the Communications Cable, part number CH0103.
T4 Standard Conguration Settings:
Alarm levels/type* H
2
S (Hydrogen Sulphide)
Low Alarm = 5 PPM
Rising alarm
Latched
High Alarm = 10 PPM
Rising alarm
STEL = 10 PPM
TWA = 5 PPM
CO (Carbon Monoxide)
Low Alarm = 30 PPM
Rising alarm
Latched
High Alarm = 100 PPM
Rising alarm
STEL = 100 PPM
TWA = 30 PPM
Rising alarm
Latched
O
2
(Oxygen)
Low Alarm = 19% Vol
Falling
Latched
High Alarm = 23.5% Vol
Rising
LEL
Low Alarm = 20%
Rising alarm
Latched
LEL (CH4)
Rising alarm
High Alarm = 40% LEL (CH
4
)
(all T4s are shipped having been calibrated with 2.2% Vol CH
4
)