54
Fault /
Warning
ID
Warning/Fault Display Image Status/Cause Operator
Action
1 “WARNING”
“ID 01”
Firmware Fault
T4 has detected an
unexpected internal
rmware fault.
Click operator
button to clear
warning.
T4 has recovered
to safe state.
T4 must be
returned to a
service centre
for further
investigation
and repair if the
message seen
persistently.
14, 15, 16,
17, 18
“WARNING”
“ID 14”
(the warning ID shown
could be one of the
codes in the adjacent
column)
Conguration
Failure
T4 has detected a
conguration read or
write failure.
Click operator
button to clear
warning.
T4 has recovered
to safe state.
T4 must be
returned to a
service centre
for further
investigation
and repair if the
message seen
persistently.
22, 23 “WARNING”
“ID 22”
(the warning ID shown
could be one of the
codes in the adjacent
column)
Logging Fault
T4 has detected it is
unable to store data
in the data or event
log.
Click operator
button to clear
warning.
T4 has recovered
to safe state.
T4 must be
returned to a
service centre
for further
investigation
and repair if the
message seen
persistently.