This determines what is output in case of an error at the
analog output.
2 mA Sensor error / System error
22 mA Sensor error / System error
None Output according Namur (3.8mA – 20.5 mA)
< 4mA to 3.8 mA Measuring range under range
>20mA to 20.5 mA Measuring range exceeding
To make changes first select a menu item "Current Error"
with button „“ and then select by pressing the „OK“
the desired mode
With „back“ change to the settings menu.