EasyManuals Logo

CS Instruments Vortex VX 570 User Manual

Default Icon
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #33 background imageLoading...
Page #33 background image
Operation
VX 570 English V1.10 Seite 33 von 44
11.3.1.5 Pressur Setup
Setup Sensor Setup Pressure Setup
Under the point "Pressure Type" the pressure
type is defined, i.e. relative or absolute
pressure.
Select the "Pressure Type" key with ">" and
change with "OK".
With "set pressure to" a pressure correction
(offset) is possible.
Select the "Set pressure to" key with ">" and
activate with "OK".
Enter / change the value by pressing the ">"
key to select the respective numerical position
and activate it by pressing the "OK" key.
Pressing ">" increases the value by 1.
Conclude with "OK" and activate the next
numerical position.
Confirm the entry by pressing the "OK" button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CS Instruments Vortex VX 570 and is the answer not in the manual?

CS Instruments Vortex VX 570 Specifications

General IconGeneral
BrandCS Instruments
ModelVortex VX 570
CategoryAccessories
LanguageEnglish

Related product manuals