D1091 Rev.M August. 2016
GENESIS V TREATMENTS WITH 1064 NM WAVELENGTH
GENESIS V TREATMENT PROCESS
• Gently heats the dermis
• 1064 nm wavelength and micro second pulses (0.3 ms)
• Preferentially heats very fine vasculature
• No epidermal cooling
TREATMENT PARAMETERS – LASER GENESIS ND:YAG NON-ABLATIVE SKIN THERAPY WITH GENESIS V HANDPIECE
These parameters are provided as a guide only for skin types I-VI. Observe laser-tissue interaction and clinical endpoints to
determine appropriate settings. Laser Genesis is recommended for scar revision, diffuse redness, fine lines, and wrinkles.
(face or neck
8 mm I – VI 4 – 7 J/cm
2
0.3 ms 7 – 10 Hz
6,000 –
8,000
4 – 6
2 – 3
weeks
(toes or
8 mm I – VI 6 – 7 J/cm
2
0.3 ms 2 – 3 Hz
Based on
area
1 – 5 4 weeks
Scars
8 mm I – VI 4 – 5 J/cm
2
0.3 ms 5 – 7 Hz
4 – 6
Temperature Monitoring
The genesis V handpiece features an integrated sensor that monitors the
temperature approximately 4 cm from the skin and displays the temperature on
the console screen. The 6 LED indicators on the handpiece correspond with
temperatures as shown. Target temperature of the skin during treatment should
be between 39
o
- 43
o
C (yellow LEDs).
LED Color Temperature Range
Red 47 – 50
o
C
Yellow 43 – 46
o
C
Yellow 39 – 42
o
C
Green 35 – 38
o
C
Green 31 – 34
o
C
o