Operator Manual

Table of Contents

Questions and Answers

  • J
    Jay MorrisAug 14, 2025
    What to do if the Cutera Excel HR Medical Equipment displays E224 during a procedure?
    • C
      Carolyn MooreAug 14, 2025
      The E224 error indicates that the handpiece cooling tip is too hot. Avoid using a thick layer of gel, as this increases the cooling requirement. Also, make sure that the heat exchanger on the bottom of the system is free of dust and debris, and ensure that the room temperature is below 27° C.
  • G
    Gloria SmithAug 17, 2025
    What to do if my Cutera Excel HR Medical Equipment system does not turn on?
    • K
      Kelsey FowlerAug 17, 2025
      First, ensure the system is properly connected to a functioning power outlet and that the main power circuit breaker is in the 'on' position. Check if the building power is on, and try resetting the room circuit breaker. If the outlet is defective, use another one or have it repaired. Also, make sure the remote interlock plug is correctly inserted into the interlock receptacle and the keyswitch is in the start position.
  • J
    Jaime KingAug 21, 2025
    Why does my Cutera Medical Equipment system shut down during operation?
    • M
      Mark HoodAug 21, 2025
      The system might shut down due to a power interruption in the building. Verify that the building power is on. Another cause could be a tripped system main power circuit breaker; if so, place it in the 'on' position. If the remote interlock is in use, ensure the treatment room door is closed, and then restart the system.
  • L
    lisa76Aug 25, 2025
    What to do if the Cutera Excel HR Medical Equipment shows a 'Water Low' advisory?
    • A
      Andrew BruceAug 25, 2025
      If the 'Water Low' advisory appears, it means the system's water level is low. Turn off the system and add water as instructed. Then, restart the system to continue treatments.
  • M
    Mary RiggsAug 28, 2025
    What to do if the Cutera Medical Equipment treatment beam and aiming beam are not present?
    • P
      Patrick PowellAug 28, 2025
      The system might be in STANDBY mode. Place the system in READY mode.
  • R
    Renee MezaSep 1, 2025
    What to do if Cutera Excel HR has inadequate aiming beam?
    • W
      Wendy JimenezSep 1, 2025
      If the aiming beam is inadequate or absent, first adjust the aiming beam intensity on the Information & Adjustment screen.
  • J
    jennifer11Sep 5, 2025
    What to do if 'Connect Footswitch' appears on Cutera Medical Equipment?
    • K
      kortizSep 5, 2025
      This message indicates that the footswitch is not properly connected to the console. Ensure that the footswitch is correctly connected. After confirming the connection, touch the (back) button in the upper left corner of the pop-up screen to clear the advisory message.
  • J
    jolsonSep 8, 2025
    Why is there low energy output in my Cutera Excel HR?
    • M
      Michele RobbinsSep 9, 2025
      Low energy output could be due to a dirty or damaged handpiece window. Confirm that the handpiece is not emitting a treatment beam using a test area. Check for the presence of the aiming beam. Clean or replace the handpiece window.

Summary

Excel HR Operator Manual Introduction

Laser Beam Characteristics

Details on the properties of the 755 nm Alexandrite and 1064 nm Nd:YAG laser beams.

System Components Overview

Identification and basic description of the main parts of the excel HR system.

Basic System Operations

Power Management and Emergency Procedures

Procedures for powering the system on/off, restarting, and critical emergency shutdown.

System Preparation and Connection

Instructions for preparing the system, connecting peripherals, and the main power cable.

User Interface and Controls

Screens, Modes, and Parameters

Navigating Select, Treatment, and Information screens, understanding modes, and adjusting parameters.

Aiming Beam and Memory Storage

Using the aiming beam for targeting and saving/recalling treatment settings.

System Maintenance and Troubleshooting

Troubleshooting and Error Codes

Guide to resolving system problems using troubleshooting steps and error code explanations.

Routine Maintenance and Cleaning

Procedures for cleaning components, handpiece, and managing coolant.

Safety and Regulatory Compliance

Optical Hazard Management and Eyewear

Critical safety information on managing optical hazards and specifying required protective eyewear.

General Safety Precautions

Precautions against electrical, fire, operational hazards, and protecting non-target tissues.

Clinical Applications and Treatment Guidelines

Indications, Contraindications, and Warnings

Summary of intended uses, conditions to avoid, and important warnings before treatment.

Treatment Guidelines by Condition

Recommended parameters and techniques for hair removal, vascular lesions, and wrinkle treatments.

Appendices

Laser Warning Sign and Symbols

Visual representation and explanation of laser warning signs and system symbols.

Cutera Excel HR Specifications

General IconGeneral
TypeLaser Hair Removal System
CoolingIntegrated contact cooling
FDA ClearedYes
Laser TypeNd:YAG
Wavelength1064 nm
Electrical Requirements50/60 Hz

Related product manuals