55
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
Operational Safety
WARNING
Laser plume may contain viable tissue particulates. The plume
presents a possible pollution hazard and should be effectively
evacuated. A commercial smoke evacuator designed for use with
aesthetic lasers may be used.
WARNING
Do not place any unnecessary stress on the umbilical cable (e.g., by
pulling on it, tightly bending it or twisting the handpiece). Doing so
may damage the cable and/or expose the patient or user to
hazardous laser radiation. See the System Specifications section
for additional information on the minimum bend radius for the
umbilical cable.
WARNING
The handpiece is a fragile instrument and must not be dropped. If it
is dropped, you must carefully examine the handpiece for any
physical damage prior to use.
WARNING
Never point the handpiece at reflective objects, such as jewelry or
smooth metal surfaces.
WARNING
Never activate the treatment beam while pointing the handpiece into
free space.
WARNING
Do not leave the system in READY mode when not in use. Always
place the system in STANDBY mode or turn off the system and
remove the key when not performing treatments.
WARNING
Do not leave the system unattended with the key in place.
CAUTION
Prior to each use of the system, inspect all protective eyewear,
cables and handpieces for any damage, excessive wear, or
crimping that could affect system performance or safe operation.