EasyManua.ls Logo

Cutera Excel HR - Electromagnetic Compatibility

Default Icon
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62
EXCEL HR OPERATOR MANUAL
D1796, REV. D, 12/16
Electromagnetic Compatibility
The excel HR laser system design complies with IEC 60601-1-2 (3rd edition) requirements for electro-
magnetic compatibility (EMC) with other devices. Like other electrical
medical equipment, the excel HR
requires special precautions to ensure EMC with other electrical medical devices and must be installed
and operated according to the EMC information provided in this manual.
Portable and mobile RF communications equipment may affect the
normal
function of the excel HR.
Do not use cables or accessories other than those provided with
the
excel HR laser system, as this may result in increased
electromagnetic emissions or decreased immunity to such
emissions.
If the excel HR laser system is used adjacent to or stacked with
other
equipment, observe and verify normal operation of the excel
HR in the configuration in which it will be used prior to using it in a
surgical procedure. Consult the tables below for guidance in placing
the excel HR.
Guidance and Manufacturers Declaration: Electromagnetic Emissions
The excel HR is intended for use in the electromagnetic environment specified below. The customer or
the user of the excel HR should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment: Guidance
RF emissions CISPR 11 Group 1 excel HR laser systems use RF energy only for
internal function; therefore, RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11 Class B excel HR laser systems are suitable for use in all
establishments, including domestic
establishments and those directly connected to
the
public low voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage Fluctuations/flicker
emissions IEC61000-3-3
Complies
CAUTION