9
TRUSCULPT OPERATOR MANUAL
D1395, REV. K, 02/18
Filling the Coolant Reservoir
The truSculpt console requires clean distilled water as a coolant. The coolant reservoir is located on the
rear of the console. Occasionally, distilled water may need to be added to top off the reservoir.
1. Turn off the system.
2. Remove the reservoir cap, and carefully fill the reservoir to the fill line using clean, unused
distilled water. Do not overfill.
Coolant Reservoir
CAUTION
Permanent damage to internal components will occur if improper
coolant is used. A container of distilled water is shipped with every
system. Under no circumstance should tap water or any liquid other
than distilled water be used.
Coolant reservoir