Error 4: Upper Display Communications issue
Cause:
Main computer board is receiving incomplete, erroneous, or no information from the upper display that
occurred even after three re-tries. The upper display is not required for the treadmill to function safely,
so this is a stored error only. If and when the product is in Dormant, the main computer board will attempt
to cycle power or reset the Upper Display system.
Service: LED Console
If this occurs in Dormant, the cause is likely a software defect or power fluctuation. If this occurs mostly
in active mode, it could be due to ESD events or increased data traffic.
Check internal cabling between the MCC and Display board.
Ensure there are no cracks or openings in the upper display’s console decal or membranes
Verify no moisture intrusion has compromised the MCC, display or cabling.
Service: EPEM (Embedded PEM or E3) Consoles
If this occurs in Dormant, the cause is likely a software defect or power fluctuation. If this occurs mostly
in active mode, it could be due to ESD events or increased data traffic.
Check internal cabling between the MCC and EPEM.
Ensure there are no cracks or openings in the upper display’s console decal or membranes
Verify no moisture intrusion has compromised the MCC, EPEM or cabling.