8-5. Actual load calibration (Exact Load Calibration)
1) Compensation Calibration
(a)Please calibrate by 8-4 Digital Calibration.
And, if there is a difference from the actual load value, you can change the indicator's
display value to the actual load value by referring to the actual load calibration setup
flow chart below.
(b) If indicator re-calibrate with 8-4. Digital calibration, all data of the actual load
calibration setting will be deleted.
(c) Change the current value to Zero (0) with key or external Zero input without
loading the sensor. After loading with a prepared standard load, change the value
displayed in the indicator to the value of the standard load by referring to the setup
sequence chart below.
Measuring mode
While
key is being
Press, press key
Display Function mode
Display Sensor output value calibration mode
Display Actual load calibration mode
After changing to actual load value using and key.
Press the
key. (500.0 is an example value arbitrarily
changed)
Actual load is applied to the sensor and the indicated value
flashes.(500.7 is an arbitrary example value)