56 Spyder2 User’s Manual
03-032-10091-06 Teledyne DALSA
Returned value Description
2
Command code: correction_calibrate_fpn
0
Er r or code: successful
192 Informal message: code 64 + code 128
33
M oni tor i ng task message: code 1 + code 32
The following tables list the codes for errors, informal messages, and monitoring task
messages. (Refer to section E7 for a list of all avail able commands.)
Error Codes
Code Description Suggested Cause
0 Command executed successful ly Command executed w i thout major error
detected
1 Internal camera error (PIXEL
IN DEX)
Internal software error trap
2
Internal camera error (RESULT
CODE)
Internal software error trap
3
Invalid command
Command not recognized
4
Command parameters incorrect
or out of range
Command parameters are invalid or out of
range
5 Command not avai lable in
current exposur e mod e
ssf
and
set
are allowed only in mode 2
(ssf and set) and i n mode 6 (set)
6 Command available in
CA LIBRA TED mode only
ccf
,
ccp
7 Command available in
UN CA LIBRA TED mode only
cao, cag
8 Command not avai lable in
VIDEO TEST mode
sao, sg
9 Start value must be an odd
number less than the even
numbered end value
roi
: invalid starting or ending pixel index
10
Camer a memory check fai l ur e
M emor y test of external RA M fai l ed
11 Unable to configure DSP DSP (FPGA) configuration failure
12
DSP configuration reset failure
DSP (FPGA) could not be placed in
configuration mode
13 Get line process command timed
out, check for the pr esence of
external signals
gl, gla, cao, cag, ccf, ccp timed out.
Current exposur e mod e r equires exter nal
SYN C, how ever SYN C signal not present
14 DSP echo test error DISC bus communication test failure
15 Invalid sensor configuration
(D SP)
Invalid sensor configuration read from DSP
(FPGA )
16 Invalid sensor configuration
(ADC)
Invalid sensor configuration encoding on
the board
17 Sensor configuration mismatch Sensor configuration read from DSP is
different than one encoded on the board
18 One (or mor e) of the supply
voltages is out of specification
vv command result
19 The camer a's temper ature is
vt
command result