4
1.3
OHMS
CALIBRATION
1.3.1
General
The
procedure
in
the
table
opposite
is
all
that
is
necessary
to
completely
'AUTOCAL'
the
Ohms
function'
lf
just
the
Ohms
or
just
one
range
of
the
Ohms
is
to be
calibrated, then
steps
1
and
2
in the
DC
Voltage
Calibration
table
should
be
carried
out first. Then
on
each
Ohms
range
just
a
'Zero'
and 'Gain'
calibration
ís
required.
lf the
'DVM
Reading
After
Calibration'
is not
in
accordance
with
the table,
repeat
operations
of
the
same
'CALIBRATE'
key
is
permissible
to
improve
the
reading'
Where no tolerance
is
shown in this
column,
only
the
exact
reading
quoted
with
an occasional
least
significant
digit
showing
is
to be
expected.
1.3.2
'Ze¡o'
Resistance
Source
For accurate'Zero'calibration
on
Ohms
it
is
ESSENTIAL
that a correctly
connected
zero source
is
used.
Two
arrangements are
necessary as shown
in
Fig.
1.1;
it
can
be seen that'4
wire Sl'selection is recommended
on
all
ranges.
1.3.3
Equipment Required
A
set of resistance standards
from 10Q
to
1OMQ
in
decades;
it is
essential that 10Q to
100kO
standards
are
4
terminal
devices. e.g. Datron
4000/40004
Autocal
Standard
with Option 20.
1.3.4
Checking
Accuracy after
'AUTOCAL'
To check
the accuracy after
'AUTOCAL'the
'Speci-
fication
Verification'
section
of the
Operating
Handbook
will
be useful.
lt
provides
tables
for
quick
reference
of
accuracy
on all ranges
and functions
in displayed
digits.
ç
1MO
and
lOMn
lOft
1o
1O0kn
I-
Lo
\
_\
I+
Hi
I+
Fis. 1.1 zERO RESISTANCE
SOURCE
CONNECTIONS
(Local
Guard
)-
O Guard
Guard
O Guard
Guard
Copper
Shorting
Link
(
Local
Guard)-
Lo
I-