3300-0192-0M Rev D 01/17 20
Setup Menu Page 5
MENU PAGE 5
Model: Icon
Calibrate Scale
<<< Exit
Model:
This will change the default settings and operation of the scale. For instance, select “Icon”
as the model in order to use the stand-up type scale with the sonar height measurement.
For wheelchair type scales with tare functionality, select “Wheelchair” as the model.
Touch this key to open the model selection screen. Using the numeric keys, enter a new
value and then press the Save key to save it.
Allowable settings are: 0 = icon, 1 = Wheelchair and 2 = Other
Calibrate Scale
Touching this key will allow for calibration of the scale. The scale display will first prompt for
the loaded calibration weight. Enter in the weight that will be used to calibrate the scale and
touch the Save key. The scale will perform the loaded calibration and then prompt for you
to unload the scale. Make sure all weight has been removed from the scale before
selecting Save to perform the unloaded calibration.
You will be returned to the menu when calibration is complete.
Press the <<< key to return to the Setup Menu Page 4.
Press the Exit key to exit Setup and return to the Weight screen.