3300-0192-0M Rev D 01/17 21
SCALE CALIBRATION
To Enter Calibration Setup
1. Enter the Setup menu of the scale by pressing the Menu key.
2. Enter in the desired setup parameters for the scale.
3. Select Calibrate Scale from the Setup Menu Page 5.
4. Make certain the scale platform is empty and free of debris.
5. Place the desired amount of calibrated test weights on the scale platform. A minimum
of 50% of scale's capacity is required. However 70% to 100% is recommended.
6. Enter the amount of the test weight into the Calibration menu screen and touch the
Save key.
7. Wait while the scale performs the loaded calibration.
8. When the loaded calibration is complete, the scale display will prompt to unload the
scale.
9. Remove any weight from the scale platform and touch the Save key.
10. Wait while the scale performs the zero calibration.
11. The scale display will return to the Setup Menu screen when the calibration is
complete.
Press the Exit key to exit Setup and return to the Weight screen.