5.5 Main Menu
Once the CL1/CL2 is configured the main menu contains all the screens necessary for normal operation. The
main menu is not password protected.
Figure 5.6
Each selection on the main menu is detailed below
5.5.1 Start Analysis
Selecting this menu item will bring up the analysis screen below and start gas analysis. Pressing <F1> during
analysis will end analysis and return to the main menu.
Figure 5.7
The screen continuously displays the current gas concentration if the system remains in analysis mode.
WARNING – Tape Saver Mode may affect the display of actual gas concentration. See section 5.5.2 for a
complete explanation of Tape Saver Mode (TSM).
When the concentration exceeds one or both alarm level text
“Alarm 1”
and
“Alarm 2”
will blink
appropriately. Anytime Alarm 1 is active the concentration display will also blink. When the concentration
passes the alarm levels the corresponding alarm output relay is triggered. The trigger will remain set
depending on the setting of Latching Relays – see section 5.4.3.