Appendix H – Advanced Optics Cleaning Procedure
NOTE: Only perform this procedure when CL1 & CL2 display reads “DIRTY” or DAC value exceeds 150.
1. Enter the CL1 Main Menu
2. Enter the Service Menu (Password: 1234)
5. Remove the ChemLogic
®
cassette tape from the gate
6. After the gate is open, remove the (2)
Phillips screws that hold the back-plane
block cover.
7. Remove the (2) Phillips screws holding the
back-plane block. The face of the optic
block should be exposed once completed.
8. You should see (4) circular holes with a
green light flashing in one channel at a time.
Each one of these holes represents a
channel. The CL1 only uses channel 2 for
gas detection