28
DC-ITD-CL1andCL2MAN01-2018.A www.dodtec.com
NOV 2018 815-788-5200
CL1 & CL2 Manual
5.7 Service Menu
The Service menu is intended for client use but should remain password protected to prevent unauthorized
access. The default Service Menu password is included with this manual. The password can be changed as
needed (see section 5.6.3)
5.7.1 Voltage Cal
This menu item is for Factory or Service personnel use, under normal conditions this screen is not used. Hit
<Cancel> to return to the Service Menu.
5.7.2 Optic Calibration
This screen is used to manually calibrate the optics which normally only needs to be done when the system is
first configured. Touch the <Start> soft key to start the calibration process. When the process starts the
DAC & Ref values will be set to 0. When the process completes both values will update with non-zero values
indicating that the process is complete.
Figure 5.27
5.7.3 Passwords
This screen is used to change the passwords for both the Setup Menu and Service menu. A password of 0
will disable the password feature for the specified menu. The default passwords are included with this
manual.
Figure 5.28
WARNING Be sure to keep the Service menu password in a safe place. If the service password is lost, you
will no longer be able to access the service menu and may need to restore default.
Use the arrow keys below the screen to select among the password fields and enter the values using the
procedure outlined in section 5.1.3. The passwords are automatically saved and set to the value displayed.
5.7.4 Optic Config
Selecting this menu item will display a screen of values that may aid service personnel in verifying the optic
system. Under normal conditions the screen is not used. Hit <ESC> to return to the Service Menu.
5.7.5 Status
Selecting this menu item will display a screen of values that may aid service personnel in finding system
faults some of which may be edited in the screen. Under normal conditions the screen is not used. Hit <ESC>
to return to the Service Menu.

Table of Contents