55
DC-ITD-CL1andCL2MAN01-2018.A www.dodtec.com
NOV 2018 815-788-5200
CL1 & CL2 Manual
16. Repeat the cleaning process until this value is back to the manufacturer's DAC value range below
130.
17. Once the value is below 130 re-install the optic block covers.
H.1 Optic Orifice Cleaning
This procedure describes the cleaning of the CL1 / Cl2 optic orifice to prevent the buildup of debris around
the orifice that could cause increased pressure readings and decreased pump voltages.
1. Press F1 to exit analysis
2. Make sure the gate is closed
3. POWER OFF the CL1 / CL2 unit
4. Carefully remove the optic exhaust tube without pulling the optic wires loose
5. Spray compressed air into the optic exhaust channel down towards the inlet of the CL1 (You should
feel air coming out of the inlet)
6. Reinstall the optic exhaust tube
7. POWER ON the CL1 / CL2 unit