Appendix D – System Event Messages
Contact DOD Technologies Support @ (815)
788-5200
See manual appendix H for cleaning
instructions. Add filter to inlet if possible. Also,
re-adjust optics - make sure fresh tape is
inserted and gate is closed. Go to service
menu-Optic Calib and press start.
Select gas in setup menu-select gas option
Make sure that all wires coming from the HMI
to the motor are fully inserted(Red (grey)=Q12
and black=0V). Verify belts are all connected
and cam is greased. If these steps do not
work, contact DOD Technologies Support @
(815) 788-5200
Motor did not return
home
Check issue from "Gate and Tape" screen. If
the gate opens and does not stop, there is a
problem with micro switch. Make sure both
wires are connected (I8 and V+). Slightly bend
the switch arm toward the motor collar to
ensure the set screw is closing the switch. If
these steps don't work, contact DOD
Technologies Support @ (815) 788-5200
Go to "Gate & Tape" and load fresh tape.
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
Also, re-adjust optics - make sure fresh tape
is inserted and gate is closed. Then go to
service menu-optics and press start.
Remove optic block cover and make sure all
wires going to the optics are fully inserted.
Contact DOD Technologies Support @ (815)
788-5200 if these steps do not correct the
problem.
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup Menu and
enable pumps. If problem persists, contact
DOD (815) 788-5200.