43
DC-ITD-CL1andCL2MAN01-2018.A www.dodtec.com
NOV 2018 815-788-5200
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup Menu and
enable pumps. If problem persists, contact
DOD (815) 788-5200.
Out of Analysis Beyond
Timer
Wrong tape or optic
problem
Re-adjust optics - make sure fresh tape is
inserted and gate is closed. Then go to service
menu-Optic Calib and press start. If these
steps do not work, contact DOD DOD
Technologies Support @ (815)788-5200
Find block in sample line
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
Also, re-adjust optics - make sure fresh tape
is inserted and gate is closed. Then go to
service menu-Optic Calib and press start.
If a SD card is installed remove and re-insert
ChemLogic tape nearing
end
Flow Blocked and/or
Unstable Inlet/Exhaust
Go to Setup Menu and enable pump. If
problem persists, contact DOD (815) 788-
5200.
Flow Blocked and/or
Unstable Inlet/Exhaust
Go to Setup Menu and enable pump. If
problem persists, contact DOD (815) 788-
5200.
Flow Blocked and/or
Unstable Inlet/Exhaust
When blockage is fixed, go to Setup Menu and
enable pumps. If problem persists, contact
DOD (815) 788-5200.
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
If problem persists, contact DOD (815) 788-
5200.
After gas leak is found and corrected, reset
alarms
After gas leak is found and corrected, reset
alarms
Pump drawing high
current or Bad Pump
Leak somewhere between the optics and
pumps. Check that pumps are connected and
all fittings between the optics and the pumps
are fully inserted. If problem persists, contact
DOD (815) 788-5200.
Pump drawing high
current or Bad Pump
Leak somewhere between the optics and
pumps. Check that pumps are connected and
all fittings between the optics and the pumps
are fully inserted. If problem persists, contact
DOD (815) 788-5200.