CL1 & CL2 Manual
47
Warning : Optics
Dirty
Dirty Optics
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
If problem persists, contact DOD (815) 788-
5200.
48
Tape Saver Mode
Tape Saver Enabled
N/A
49
Hour Diagnostic OK
Hourly Checkup
N/A
50
Flow Failure 1 L
Optics Orifice Dirty
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
If problem persists, contact DOD (815) 788-
5200.
51
Flow Failure 2 L
Optics Orifice Dirty
See manual appendix H for cleaning
instructions and add a filter to inlet if possible.
If problem persists, contact DOD (815) 788-
5200.
52
Flow Disruption
Flow Inconsistent
Contact DOD Technologies Support @ (815)-
788-5200
53
Alarm Level 1 Pt 2
Gas Detected
After gas leak is found and corrected reset
alarms
54
Alarm Level 2 Pt 2
Gas Detected
After gas leak is found and corrected reset
alarms
55
Flow Failure 1 H
Sample Inlet Blocked or
Internal Leak
Leak somewhere between the optics and
pumps. Check that pumps are connected and
all fittings between optics and pumps are fully
inserted. Also, make sure all three wires are
connected to the transducer. If problem
persists, contact DOD (815) 788-5200.
56
Flow Failure 2 H
Sample Inlet Blocked or
Internal Leak
Leak somewhere between the optics and
pumps. Check that pumps are connected and
all fittings between optics and pumps are fully
inserted. Also, make sure all three wires are
connected to the transducer. If problem
persists, contact DOD (815) 788-5200.