5.6.4 Enable Pumps
(Default Configuration: Enabled)
Press the upper soft keys to toggle between enabled/disabled on the corresponding pump. The display
adjacent to the soft key always displays the current state of the pump. (ex: in figure 5.18 both pumps are
currently enabled). When a pump is disabled a flashing ‘P1’ or ‘P2’ corresponding to the disabled pump will
appear in the lower left corner of the analysis screen. (See section 5.4.2) Disabling both pumps will trigger a
critical system fault and prevent analysis mode.
Figure 5.18
5.6.5 Enable Points
(Only Available on CL2)
Like the Enable Pumps screen, this screen will allow either point on a CL2 to be enabled/disabled.
5.6.6 Energized Alarm Relays
(Default Configuration: OFF)
Use the upper left soft key to select Energize/De-Energized alarm relays. The soft key always shows the
CURRENT state of the relays the initial default value is Energized.
Figure 5.19
5.6.7 Idle Timeout
(Default Configuration: 45 Minutes)
Enter the number of minutes before an ‘Idle Timeout Fault’ occurs. When the system is out of analysis for the
specified amount of time the fault will occur. Specifying zero (0) minutes will disable the fault from occurring.
The maximum amount of time allowed is 45 minutes.
Figure 5.20