9. To clean the fiber optic lenses, you will need
a Small cotton swab(Q-tip) and Industrial
alcohol.
10. Dip the cotton swab into the alcohol and
insert it in channel 2. It is very important to
clean both lenses. Make sure to put the
swab straight in to clean the first lens and
angled downward to clean the second lens.
11. After the cleaning is complete re-insert the tape and put the back-plane block on. Press
"Open/Close" and the gate will close.
12. Press start to initiate the calibration process. The current DAC value will become 0. Then the new
DAC value will populate. This new value should between 60-130 which is the manufacturers range.
DAC value cannot be improved from the value it is shipped at. Call DOD Technologies to check your
starting DAC value.
13. When done, the main menu should be on the display and press start analysis.
14. Once in analysis press F6 in order to see the technician screen.
15. In the technician screen you will see the
letter D. In front of the D is a number that
represents how bright your optic LED is.
When this number reaches 140 you will be
prompted to clean with a fault. When this
value reaches 150 your CL1 will not go in to
analysis until the optic block is cleaned.