3736
P750 (1-channel) / P755 (2-channel)
Inputs 1/2 Pt100,
thermocouple: Typ K, J, L, N, R, S, T
humidity, ow, resistance, voltage
Measuring range
Pt100 -200...+850 °C
Thermocouple according DIN (-200...1760°C)
Resistance 0 ... 400 Ω
Humidity 0 %...100 %rF
Flow 0 ... 40 m/s
Accuracy
Pt100 ±0.03°C from -50°C...+199.99°C
±0.05°C from -200°C...-50.01°C
otherwise 0.05%
Thermocouple
R, S ±1.0 °C +0.1%
K, J, L, N, T ±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Humidity ±1.5%rH
Flow 1% of end value
Resistance 0.5%
Resolution 0.0 1°C from -200 °C...+200 °C,
otherwise 0.1°C, respectively 0.1%
Ex-mark -
Battery life appr. 20 hours
Connectors DIN 8-pole
Allowable operating temp. 0 °C ... +40 °C
Display 2-line LCD
Housing plastic
Dimensions 200 x 93 x 44 mm (LxWxH)
Weight 350 g
Power supply 9 V battery
P770/ P770B (2-channel):
Inputs 1/2 Pt100,
thermocouple: Typ K, J, L, N, R, S, T
humidity, ow
Measuring range
Pt100 -200...+850 °C
Thermocouple gem. DIN (-200...1760°C)
Resistance -
Humidity 0 %...100 %rH
Flow 0 ... 40 m/s
Accuracy
Pt100 ±0.1 °C from -100 °C...+200 °C
otherwise 0.1%
Thermocouple gem. DIN (-200...1760°C)
R, S ±1.0 °C +0.1%
K, J, L, N, T ±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Humidity ±1.5%rH
Flow 1% of end value
Resistance -
Resolution 0.1°C, 0.1% and 0.1 m/s
Ex-mark -
Battery life appr. 20 hours
Connectors DIN 8-pole
Allowable operating temp. 0 °C ... +40 °C
Display 2-line LCD
Housing plastic
Dimensions 200 x 93 x 44 mm (LxWxH)
Weight 350 g
Power supply 9 V battery
P755-LOG/ P755B-LOG (2-channel)
Inputs 1/2 Pt100,
thermocouple: Typ K, J, L, N, R, S, T
humidity, ow, resistance, voltage
Measuring range
Pt100 -200...+850 °C
Thermocouple according DIN (-200...1760°C)
Resistance 0 ... 400 Ω
Humidity 0 %...100 %rF
Flow 0 ... 40 m/s
Accuracy
Pt100 ±0.03°C from -50°C...+199.99°C
±0.05°C from -200°C...-50.01°C
otherwise 0.05%
Thermocouple
R, S ±1.0 °C +0.1%
K, J, L, N, T ±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Humidity ±1.5%rH
Flow 1% of end value
Resistance 0.5%
Memory appr. 6,000 measurements
Resolution 0.0 1°C from -200 °C...+200 °C,
otherwise 0.1°C, respectively 0.1%
Ex-mark -
Battery life appr. 20 hours
Connectors DIN 8-pole
Allowable operating temp. 0 °C ... +40 °C
Display 2-line LCD
Housing plastic
Dimensions 200 x 93 x 44 mm (LxWxH)
Weight 350 g
Power supply 9 V battery
P770 (2-channel):
Inputs 1/2 Pt100,
thermocouple: Typ K, J, L, N, R, S, T
humidity, ow
Measuring range
Pt100 -200...+850 °C
Thermocouple gem. DIN (-200...1760°C)
Resistance -
Humidity 0 %...100 %rH
Flow 0 ... 40 m/s