Notice:
Defective or empty batteries can lead to measurement errors or error messages
produce. In the event of error messages, remove the battery and wait approx.
3 minutes before inserting a new battery.
3. Error Codes
Important! By displaying the following error codes the instrument support the
operation of the instrument.
Error Meaning
oPEn no probe or wrong probe is connected
“too low“ – underow of the measuring range
„too high“ – exceeding of the measuring range
temperature of the cold junction above the measuring range
temperature of the cold junction above the measuring range
E15 low battery – (battery is com-pletely exhausted)
E19 low battery – (battery is com-pletely exhausted)
E16 time out / auto-off-function
E1 time out / auto-off-function
E1e time out / auto-off-function
E23 EE-prom destroyed or EE-prom index is wrong
E25 EE-prom destroyed or EE-prom index is wrong
E31 temperature of cold junction out of range
LibA Internal buffer battery (CR2032) empty
(for changing the battery the housing must be unscrewed)
r
E
r1
maximum statistical value cannot be calculated
r
E
r2
maximum statistical value cannot be calculated
4. Technical data
P700/P700B (1-channel) / P705 (2-channel):
Inputs 1/2 Pt100, thermocouple: Typ K, J, L, N, R, S, T
Measuring range
Pt100 -200...+850 °C
Thermocouple according DIN (-200...1760°C)
Accuracy
Pt100 ±0,1 °C from -100 °C...+200 °C 0,1% remaining range
Thermocouple
R, S ±1,0 °C +0,1%
K, J, L, N, T ±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Resolution 0,1 °C
Ex-mark -
Battery life appr. 20 hours
Connectors DIN 8-pole
Allow. operating temp. 0 °C ... +40 °C
Display 2-line LCD
Housing plastic
Dimensions 200 x 93 x 44 mm (LxWxH)
Weight 350 g
Power supply 9 V battery
P710 (1-channel) / P715 (2-channel)
Inputs 1/2 thermocouple: Typ K, J, L, N, R, S, T
Measuring range
Pt100 -200...+850 °C
Thermocouple according DIN(-200...1760°C)
Accuracy
Pt100 -
Thermocouple
R, S -
K, J, L, N, T ±0.2 °C from 0 °C...+200 °C
±0.5 °C to 1000 °C
±1.0°C remaining range
Resolution 0.1°C
Ex-mark -
Battery life appr. 20 hours
Connectors DIN 8-polig
Allowable operating temp. 0 °C ... +40 °C
Display 2-line LCD
Housing plastic
Dimensions 200 x 93 x 44 mm (LxWxH)
Weight 350 g
Power supply 9 V battery

Table of Contents

Other manuals for Dostmann Electronic P700 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dostmann Electronic P700 Series and is the answer not in the manual?

Dostmann Electronic P700 Series Specifications

General IconGeneral
BrandDostmann Electronic
ModelP700 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals