Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 105
Operation
An insufficiently filled breathing bag can trigger the Fresh gas low or leakage or
Emergency air inlet activated alarms, for example.
7.5.3 Stopwatch
Using the stopwatch
1. Touch the Start button (1) to start.
2. Touch the Stop button (1) to stop.
3. To reset the stopwatch to zero, touch the Reset button (1).
7.5.4 Timer
Setting the timer
1. Touch the Set button (1) or the parameter field.
2. Set the timer.
Using the timer
The timer always starts with the last time set.
1. Touch the Start button (1) to start.
2. Touch the Stop button (1) to stop.
3. To reset the timer to zero, touch the Reset button (1).
Range Color Meaning
Surplus Yellow Indication of an opportunity to save fresh
gas and, therefore, volatile anesthetic
agents
Efficient Green No action necessary
Breathing bag sufficiently filled
Sufficient reserve capacity available
Deficit Red Insufficient fresh-gas supply
Check the filling of the breathing bag. If
necessary, fill up the breathing bag,
e.g., with the O
2 flush.