272 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Technical data
Pressure-related settings
Inspiratory pressure Pinsp PEEP +5 to 80 hPa (cmH
2O)
±10 % of the set value or ±3 hPa (cmH
2O)
(the larger value applies)
Pressure limitation Pmax PEEP +10 to 80 hPa (cmH
2O)
±10 % of the set value or ±3 hPa (cmH
2O)
(the larger value applies)
Relative pressure support above PEEP Psupp Off, 3 to (80 - PEEP) hPa (cmH
2O)
±10 % of the set value or ±3 hPa (cmH
2O)
(the larger value applies)
in the CPAP / PSV and PC - SIMV / PS modes
Off, 3 to (Pmax - PEEP) hPa (cmH
2O)
±10 % of the set value or ±3 hPa (cmH
2O)
(the larger value applies)
in the VC - SIMV / PS and VC - SIMV / PS /
AutoFlow modes
Positive end-expiratory pressure PEEP Off, 2 to 35 hPa (cmH
2O)
±10 % of the set value or ±2 hPa (cmH
2O)
(the larger value applies)
Minimum pressure limit as per ISO 8835-5 and
ISO 80601-2-13
-3 hPa (cmH
2O)
16.8 Breathing system
Total volume (when applying a maximum
tidal volume of 1500 mL)
with CLIC absorber 800+ disposable CO
2
absorber
3.65 L ± 150 mL
with reusable CO
2 absorber and Drägersorb
800+
3.83 L ± 150 mL
Compliance (without breathing circuit)
In the Man / Spon mode (including disposable
CO
2 absorber)
Typically 2.7 mL/hPa (mL/cmH
2O)
corresponds to 81 mL at 30 hPa (cmH
2O)
for the "Neonates" patient category
Typically 2.8 mL/hPa (mL/cmH
2O)
corresponds to 84 mL at 30 hPa (cmH
2O)
for the "Pediatric patients" patient category
Typically 3.0 mL/hPa (mL/cmH
2O)
corresponds to 90 mL at 30 hPa (cmH
2O)
for the "Adults" patient category
16.7 Ventilator (continued)