Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 311
Principles of operation
17.6 Influence of patient category, weight, and age on
device behavior
17.6.1 Influence of patient category
Alarm limits and start settings for therapy
Volumeter scale
Flow measurement and software algorithms to suppress artifacts
Maximum duration of a pressure-supported breath
17.6.2 Influence of ideal body weight and height
Ideal body weight describes the portion of the body that is relevant to setting the
ventilation parameters (body weight of the patient minus assumed excess fat).
In the Adult and Ped patient categories, the ideal body weight is calculated from
the entered body height.
In the Neo patient category, the ideal body weight is equal to the entered body
weight.
The calculated ideal body weight affects the following:
Start settings for tidal volume VT
Start settings for respiratory rate RR
Start settings for alarm limits for VT and MV
Ventilation algorithms
Trigger threshold (insp. synchronization of the mandatory breaths)
Sensitivity and resolution of the volume monitoring
Rate of concentration changes affecting the patient
For neonates, the patient category and the ideal body weight must be set with
particular care.
VT and RR are only dependent on the ideal body weight when the Ideal body
weight function has been selected in System setup > Therapy, see page 172.
Mechanically
controlled
21 % 21 % Continuously adjustable with flow
control valves.
The minimum O
2 delivery interrupts
the N
2O flow in the following cases:
–N
2O flow control valve open and
O
2 flow control valve closed
–O
2 flow less than 200 mL/min
O
2 will continue to be delivered in
the event of an N
2O failure.
Gas mixer Minimum FG O
2
concentration
Minimum O2 flow
Carrier gas
Air
Carrier gas
N
2O