EasyManuals Logo

Dräger Atlan A350 XL User Manual

Dräger Atlan A350 XL
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
242 Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n
Reprocessing
On the bottom side of the lower part of the breathing system:
Turn the black fresh-gas decoupling valve (3) counterclockwise and remove it.
13.9.2.6 List of disassembled components
After complete disassembly of the breathing system, the following machine-
reprocessable components are present:
Upper part of the breathing system
Lower part of the breathing system
Sensor cap of the O
2 sensor
–APL valve
Inspiratory port
Expiratory port
Inspiratory valve (yellow)
Expiratory valve (blue)
Fresh-gas decoupling valve (black)
Piston diaphragm
Breathing bag arm (option)
Optionally present are:
Absorber container and absorber insert
CLIC adapter (option)
13.10 Machine reprocessing
13.10.1 Preparation
Prerequisite: The breathing system and components are completely disassembled.
Information on the number of reprocessing cycles
At least 250 reprocessing cycles are possible for the breathing system.
37060

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 XL and is the answer not in the manual?

Dräger Atlan A350 XL Specifications

General IconGeneral
BrandDräger
ModelAtlan A350 XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals