Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 163
Alarms
10.6 Activation of alarms after breath detection
Product variants with inspiratory O2 measurement or integrated patient-gas
measurement module
In the Man / Spon mode, the alarms for Minute volume low and Apnea (no flow)
are only activated after spontaneous breaths have been detected.
Product variants with integrated patient-gas measurement module
If still no breaths have been detected after leaving the Standby or Pause modes,
the breathing gas is monitored with regard to an O
2 concentration that is too low or
an anesthetic gas concentration that is too high. At the same time, the message
Waiting for respiratory phases is displayed in the CO
2 waveform.
Once 2 breaths have been detected, the message disappears and only then are the
O
2, CO2, N2O, and anesthetic gas alarms active.
Cardiac bypass mode
still active?
If a minute volume of
>50 % of the suggested
value is measured after
CBM mode has been
activated for
>60 seconds
--- ---
Fresh gas low or leak-
age
--- Breathing bag empty With an additional
"Apnea (no pressure)"
alarm and at least one of
the following alarms:
"Apnea (no flow)", "Tidal
volume not achieved" or
"Airway pressure not
achieved"
Alarm Priority Priority Priority
Note
(Low)
Caution
(Medium)
Warning
(High)