EasyManuals Logo

Dräger Atlan A350 XL User Manual

Dräger Atlan A350 XL
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 145
Tests
8.4.5.3 Manual resuscitator
1. Make sure there is a manual resuscitator on the device.
2. Check the functional integrity of the manual resuscitator.
8.4.5.4 Anesthetic gas receiving system
With active anesthetic gas scavenging:
Have the flow for the anesthetic gas scavenging system set so that the red flow
indicator (1) floats in range (2) ("normal range").
If the flow indicator (1) is floating in range (3) ("restricted range"), certain fresh-
gas flows should not be exceeded, see "Anesthetic gas receiving system" in
chapter "Technical data". Contamination of the ambient air can be prevented by
limiting the fresh-gas flow.
3512135125
2
3
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 XL and is the answer not in the manual?

Dräger Atlan A350 XL Specifications

General IconGeneral
BrandDräger
ModelAtlan A350 XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals