Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 231
Reprocessing
Pull out and dispose of the water trap (1).
13.8.3 Disassembling the CO2 absorber
13.8.3.1 Reusable CO2 absorber
1. Unscrew the CO
2 absorber (1) from the breathing system.
2. Remove and dispose of the optional disposable dust filter (2).
3. CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
Handle the soda lime carefully and do not spill it.
Empty out the used soda lime and dispose of it according to its instructions for
use.
3962937107