Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 237
Reprocessing
WARNING
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Electric shock
Device malfunctions
Ensure that no liquid penetrates the device.
Perform the following steps:
1. Use a cloth soaked in disinfectant or use a ready-to-use disposable cloth.
2. Perform the surface disinfection of the breathing system mount by wiping
backwards and forwards at least 3 times. Take particular care to disinfect the
pneumatic connections (1).
3. Repeat steps 1 and 2 (4 times each).
39534