Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 247
Reprocessing
Absorber container
Place the absorber container with the opening facing down on an injector
nozzle.
CLIC adapter (option)
Lay the CLIC adapter in the basket with the cleaning plate facing upwards.
13.10.3 Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the requirements of the standard ISO 15883.
Dräger recommends the use of a load carrier for anesthesia accessories and
ventilation accessories.
Follow the list of validated disinfectants (see "Validated reprocessing procedures",
page 225).
Prerequisite: The components have been positioned appropriately in the load
carrier.
1. Use a suitable cleaning agent.
2. Select a suitable cycle.
Use a cycle with the following phases:
Pre-wash
Cleaning using a suitable cleaning agent in accordance with the
manufacturer’s instructions for use
Final rinsing with demineralized water
3859538594