Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 277
Technical data
T0...90 <45 s (RR 6/min)
<105 s (RR <6 /min)
Leakage minute volume MVleak
Range 0 to 40 L/min
Accuracy ±25 % of measured value or ±0.2 mL/min
(the larger value applies)
Resolution of displayed value 0.01 L/min (MV <1 L/min) or 0.1 L/min
(MV 1L/min)
Low-flow wizard
Standardization The following data are standardized to STPD
conditions.
Range 0 to 8 L/min
Accuracy ±25 % of measured value or ±100 mL/min
(the larger value applies)
Respiratory rate
Total RR
Spontaneous RRspon
Mandatory RRmand
Range 0 to 100 /min
Accuracy ±10 % or ±1 /min (the larger value applies)
Resolution of displayed value 1 /min
T
0...90 <45 s (RR 6/min)
<105 s (RR <6 /min)
Dynamic compliance
Dynamic compliance Cdyn
Mean dynamic compliance Cdyn mean
Range 0 to 200 mL/hPa (mL/cmH
2O)
Accuracy for gas measurement with PGM ±15 % or ±1 mL/hPa (mL/cmH
2O) (the larger
value applies)
Accuracy without PGM ±40 % or ±3 mL/hPa (mL/cmH
2O) (the larger
value applies)
Information The values may be distorted by spontaneous
breathing, thus reducing the measurement
accuracy.
Resolution of displayed value 0.1 mL/hPa (mL/cmH
2O)
Resistance R
Range 0 to 100 hPa/L/s (cmH
2O/L/s)
Accuracy for gas measurement with PGM ±30 % of the measured value or ±3 hPa/L/s
(cmH
2O/L/s) (the larger value applies)
16.13 Measuring systems and displays (continued)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Atlan A350 XL and is the answer not in the manual?

Dräger Atlan A350 XL Specifications

General IconGeneral
BrandDräger
ModelAtlan A350 XL
CategoryMedical Equipment
LanguageEnglish

Related product manuals