Instructions for use | Atlan A300, A300 XL, A350, A350 XL SW 1.0n 295
Technical data
Typical data volume:
Update of device firmware: Typically 50 MB
Help ticket (system logbook for service purposes): Typically 3 MB
While the service functions are being used, the device can cause network loading
up to the maximum transmission speed of the LAN interface. The bandwidth used
during normal use is negligibly low.
16.25.1.4 Hazardous situations
The following hazardous situations may occur if the network does not possess the
required characteristics:
Exported patient-related data (age, weight, height) and therapy-related data may
be intercepted, falsified, or damaged.
An overload of the device due to high network loading (e.g., caused by denial-of-
service attacks) may lead to a shut-down of the device-side network interface.
The interface will not be available again until the device is restarted.
16.25.2 RS-232 ports
The RS-232 port supports the MEDIBUS.X protocol. MEDIBUS.X is a
communication protocol for data exchange between the device and, e.g., the
following external medical or non-medical equipment:
Hemodynamic monitor
Data management system
–Computer
RS232 to Ethernet converter
Connections between Atlan and corresponding destinations
Function Proto-
col
Atlan
port
Direction Remote
port
Remote
partner
Connec-
tion
SNMP V3 UDP 161 >1023 SCG
TCP SNMP
V3
(Trap)
UDP >1023 162 SCG
FTP
(command)
TCP >1023 21 SCG New,
estab-
lished
FTP
(command)
TCP >1023 21 SCG Estab-
lished
FTP
(data)
TCP >1023 >1023 SCG New,
estab-
lished
FTP
(data)
TCP >1023 >1023 SCG Estab-
lished
SNTP UDP >1023 123 NTP server
DHCP UDP 67 67 DHCP
server