6173.3 Babylog 8000/8000 plus/8000 SC 12/00 General Page 9
For internal use only. Copyright reserved.
GBK61733XXTL1.fm 07.07.05
9.2 Test 2 With Control Pressure
Use the same test set-up as in Test 1 and proceed as follows:
Remove the bottom plate from the Babylog 8000/Babylog 8000 plus/
Babylog 8000 SC and disconnect the hose from connection 4 on the gas connection
block. Apply a pressure of 1.5 bar ±0.1 bar into this hose (see Figure below).
Fig. 4: Test 2
4x1.5 Si nf hose
Babylog hose disconnected from
connection
4 on the gas
connection block.
PEEP/PIP-
Valve
Valve block
Insp. block
Gas connection block
4
4
View: Gas connection block tilted by 90°