44 Instructions for use BiliLux
Operation
8.1.2 Positioning the BiliLux with spring arm and BiliLux with trolley
at an incubator
1 If the device is equipped with a trolley, adjust the height of the trolley as desired
using the adjustment knob. If equipped with spring arm without trolley, proceed
to step 4.
2 Roll the trolley under the incubator. Keep the distance between the trolley and
the incubator to a minimum.
3 Lock the castors to prevent movement of the trolley.
CAUTION
Risk of injury
Cables or attached accessories may be inadvertently disconnected or otherwise
compromised when adjusting the spring arm.
Take care not to compromise cables or accessories when adjusting the spring
arm to the desired position.
4 Adjust the spring arm to the desired height and position.
CAUTION
Risk of injury
The LED panel on the phototherapy light may be hot to the touch.
Adjust the phototherapy light only by touching the housing. Do not touch the
LED panel.
5 Position the phototherapy light over the incubator hood. Ensure that as much of
the patient’s skin as possible is covered with phototherapy light.
038
≥30 cm
(12 in)
≥30 cm
(12 in)

Table of Contents

Other manuals for Dräger BiliLux

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger BiliLux and is the answer not in the manual?

Dräger BiliLux Specifications

General IconGeneral
BrandDräger
ModelBiliLux
CategoryMedical Equipment
LanguageEnglish

Related product manuals