68 Instructions for use BiliLux
Service
12.6.3 Tightening the spring arm joints
1 Position the device so that the spring arm is in a horizontal position.
2 Remove the covers (A) on the spring arm joints.
3 Tighten the hex nut (B) in each joint using a 13 mm socket wrench so that the
device stays securely in the horizontal position. Do not overtighten.
4 Verify that the spring arm moves freely and remains in any position.
5 Reattach the covers on the spring arm joints.
12.7 Repair
Dräger recommends that all repairs are performed by DrägerService and that only
authentic Dräger repair parts are used.